ASTM D2982 - Detecting Glycol-Base Antifreeze in Used Lubricating Oils

Significance and Use

5.1 Leakage of glycol-base antifreeze into the crankcase is serious because the coolant tends to interfere with the lubricant and its ability to lubricate; it also promotes sludging. Ethylene glycol present in the coolant can increase varnish deposit formation in the crankcase as a result of glycol oxidation and the interaction between glycol and lubricant. Furthermore, because glycol is a higher boiling material than water, it will tend to stay longer in the crankcase oil than water. Lubricant displacement, sludging, and deposit formation all lead to engine malfunction and possible seizure.

5.2 These tests are designed to detect glycol-base coolant contamination even at low levels because early detection enables corrective measures to be taken to prevent leaking coolant from accumulating and seriously damaging the engine.

5.3 These test methods are also significant because the reagents can be packaged as a field kit, and the procedure can be followed at the site where there is a concern.

Scope

1.1 These test methods cover the qualitative determination of glycol-base antifreeze in used lubricating oils (mineral base) by two procedures, one using reagents in tablet form and the other using laboratory shelf reagents. Principally the test methods detect ethylene glycol but will also detect other 1,2-glycols that may be present.

1.1.1 When a positive result is obtained and a sample of the unused oil is available, the unused oil is also tested and used as a reference.

NOTE 1—Since the inception of this test method (1971), there have been many changes in base stock technology and additive technology. Therefore, when available, the new, unused oil, or a sample of the same used oil, known to not contain antifreeze, is tested as a reference.

1.2 The tablet procedure (Procedure A) is sensitive to about 100 mg/kg and the shelf reagent procedure (Procedure B) to about 300 mg/kg of ethylene glycol.

1.3 Glycol-based coolant leaks into crankcases may not be detected or may result in a low bias using these test methods if the glycol has degraded or been thermally or otherwise oxidized. The conditions in crankcases may be such that contaminant glycols are oxidized or degraded to a degree to which the color indicator reaction does not occur or is biased enough so as to not trigger the color change. Other test methods for the detection of coolants or coolant additives in lubricating oils should be used if the results from these test methods alone are inconclusive or questionable.

1.4 Carbohydrates such as sugars and sugar-containing substances are sometimes used for sabotage purposes. If the presence of these substances is suspected, Procedure A contains a modification to remove these interferences.

1.5 Both procedures are adaptable to field kit use, and brief descriptions for converting to field kit form are given in Annex A1.

Extracted, with permission, from ASTM D2982-7 - Detecting Glycol-Base Antifreeze in Used Lubricating Oils, copyright ASTM International, 100 Barr Harbor Drive, West Conshohocken, PA 19428. A copy of the complete standard may be purchased from ASTM International, astm.org